Optimasi Screen Printed Carbon Electrode Modifikasi Nikel-Carbon Untuk Analisa Glukosa Menggunakan Pudelstat Secara Cyclic Voltammetry
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Fakultas Matematika dan Ilmu Pengetahuan Alam
Abstract
The high cost of commercial potentiostats has encouraged the development
of low-cost open-source electrochemical instruments such as PUDELstat. This
study aimed to optimize a nickel-carbon ink modified Screen-Printed Carbon
Electrode (SPCE) for non-enzymatic glucose detection using Cyclic Voltammetry
(CV) with a PUDELstat potentiostat. The effects of electrode surface area (4, 5,
and 6 mm²), scan rate (50–250 mV/s), and NaOH electrolyte pH (12.0–13.5) on
glucose oxidation were systematically investigated. The analytical performance of
the optimized sensor was further evaluated using an artificial sweat matrix. The
PUDELstat demonstrated satisfactory calibration performance with coefficients of
determination (R²) of 0.96918 and 0.96264 for voltage and current measurements,
respectively. The optimum electrode surface area was 6 mm², providing the highest
oxidation peak current and good repeatability with %RSD values below 5%.
Increasing the scan rate from 50 to 250 mV/s enhanced the oxidation peak current
from 0.7125 to 0.9705 µA, indicating a diffusion-controlled electrochemical
process consistent with the Randles–Sevcik equation. The optimum electrolyte
condition was achieved at pH 13.5, where the highest oxidation current (1.198 µA)
was obtained due to the enhanced formation of the electroactive Ni(OH)₂/NiOOH
redox couple responsible for glucose oxidation. The optimized sensor exhibited
excellent analytical performance with a linear regression equation of y = 0.0621x
+ 0.024 and a coefficient of determination (R²) of 0.9985. The sensor achieved a
limit of detection (LOD) of 11,6 mM and a limit of quantification (LOQ) of 13.2mM.
Recovery testing in an artificial sweat matrix yielded a recovery value of 97% with
%RSD below 5%, indicating good accuracy and precision. These results
demonstrate that the nickel-carbon ink modified SPCE integrated with PUDELstat
is a reliable, accurate, and low-cost platform for non-invasive glucose monitoring
in sweat matrices.
Description
:: Finalisasi file repositori 21 Juli 2026_Kurnadi